2-2 Weight Calibration

In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to
enter the External Function setting mode. The LED shows
NET/GROSS or UNITS key to select
ACTIONS
Press TARE key
When the digit on the right
flashes, ensure that the
weigh pan is empty. Then
Press the TARE
key.
After stabilization
Place the displayed weight
on the weigh pan. Press the
TARE key. The scale
calculates the full capacity
internal value and after
stabilization, the buzzer
beeps 3 times. Remove the
weight from the weigh pan
and the procedure
is completed.
For approved models,
Weight calibration conditions:
The calibration weight value placed on the weight pan must be over e100, and the
standard deviation of the weight must be within 10%.
02005324
DISPLAY
.
.
is disabled.
17
to enter the weight calibration mode.
NOTE
The LED displays the calibration
weight value (the maximum capacity
st
of 1
unit) and the unit. The digit on
the right side keeps flashing.
The scale is calculating the internal
value of the zero point. After
stabilization, the LED shows the
calibration weight value.
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
and uses the
ZSME300000255