1-4 Self-Test Mode

Set the switch SWA1 on the bottom of machine to the LOCK position.When the scale
count down, press NET/GROSS and ON/ZERO key together, Wait till display shows
to enter "Self-Test Mode".
Press
ON/ZERO
INTERNAL VALUE MODE
Press TARE to enter, and the display shows internal value
Please check whether the internal value has changed obviously with weight changing.
Please check the backlight.
Press ON/ZERO key to back to the last screen , the display shows
KEYPAD TEST MODE
Press TARE to enter, display shows
Keypad's internal code: TARE key= 07 、 UNITS key= 06、 NET/GROSS key=05
Press ON/ZERO key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
Press TARE to enter, display shows the firmware version
Press TARE key again, the display shows maintenance number
Press ON/ZERO key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press TARE key to exit self-test mode, the scale will restart automatically.