Limitations of Blood glucose measuring system :
The blood glucose test Strips provide accurate results when the following constraints are
observed :
- Use the
blood glucose test strips with the
test meter or
blood glucose test meter.
- Use fresh capillary whole blood only. Do not use plasma or serum
- Do not use neonate samples.
- The blood glucose test strips are for single use only. Do not reuse.
- Dehydration may lower test results. If you are severely dehydrated, contact
you physician immediately
- Inaccurate results may occur when in shock, hypotensive, hyperglycemic,
or hyperosmolar state, with or without ketosis.
- Blood glucose test strips used above altitudes of 9,840 feet will have an
effect on test results.
Doctors - Please note the following interferences that may affect test results :
-
test strip results can be used with hematocrit levels in the range of 20
to 60
. If the hematocrit range is out of the range(20%~60%), then the test results
%
may be lower or higher than actual value. If you do not know your hematocrit level,
consult your healthcare professionals.
- Interferences : Acetaminophen, Uric and, Ascorbic acid (Vitamin C), and other
reducing substances when occurring in normal blood or normal therapeutic
concentrations do not significantly affect results. However, abnormally high
concentrations in may cause inaccurately high results.
- Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 2890 mg/dl
do not significantly affect the results. Values beyond these levels should be
interpreted with caution.
- Blood samples that contain a high concentration of dissolved oxygen may
lower the test result.
Performance Characteristics :
The performance of the blood glucose test strips has been evaluated in
laboratory and in clinical tests.
Measurement Range : The measurement range of the
glucose measuring system is 10 to 600 mg/dL (0.6~33.3mmol/L).
12
Before You Test
The
functions to choose from. There is a beeping sound to alert you when a function
is set; date, time, five daily alarms, unique user ID, ability to designate activities
blood glucose
related to the test result, and the ability to designate three unique data averages.
Turn on
Press power button
and hold for
%
about 3 seconds
User's ID
After pressing and
releasing either
buttons to turn user ID
Blood
on/off, enter
Blood Glucose Monitoring System has a wide variety of
Beeping sound
After pressing and
releasing either
or
buttons to turn the
beeper on/off, enter
Date/Time
After pressing and releasing
either
or
buttons to set
or
the date and time, enter
User's testing option
After pressing and
releasing either
or
buttons to turn the
user testing option
on/off, enter
Test unit
After pressing and
releasing either
or
buttons to set the test unit
of result, enter