Trouble Shooting
MESSAGE
Problem
Problem with the test strip
Test strip is damaged.
User applied blood sample
before the LCD display,
blinking.
No readings in memory to display
Not enough readings in memory
to display designated averages
38
Action Required
Please insert a new test strip
and perform your test again.
Please wait for the LCD to
display,
blinking before
applying your blood sample.
Trouble Shooting
MESSAGE
Problem
The ambient temperature
is less than 10°C (50°F)
The ambient temperature is
over 40°C (104°F)
The test result is lower than
10 mg/dl or 0.6 mmol/L
The test result is higher than
600 mg/dl or 33.3 mmol/L
Action Required
Place the meter within the
operating temperature range
for more than 30 minutes
and retest.
Place the meter within the
operating temperature range
for more than 30 minutes
and retest.
Check the accuracy of your test
strip by perfoming a control
solution test. With normal result,
re-test with blood sample two or
three times. When "Lo" message
persists, please consult a doctor
for assistance.
Check the accuracy of your
test strip by perfoming a control
solution test. With normal
result, re-test with blood
sample two or three times.
When "HI" message persists,
please consult a doctor for
assistance.