You can set a temporary basal rate for a duration of 30 minutes
to 12 hours. Once the time limit is reached, the PDM returns to
the current basal program.
When using units per hour, the OmniPod System allows you to
set temporary basal rates from 0.0 U/hr, up to your max basal
rate. When using percent rate, you can set temporary basal rates
from -100% to +95% of your current rate, as long as the tempo-
rary rate is less than or equal to your max basal rate.
See the sections below on creating, enabling, canceling, and
changing temporary basal presets.
Enable a one-time temporary basal rate
1. On the Home screen, use the Up/Down Controller button
to choose Temp basal; then press Select.
2. Choose [enter manually] from the list of temp basal presets,
then press Select. (If you have not created any temp basal
presets, the PDM skips this step.)
3. Enter the temporary basal rate in either units per hour (Figure
3-10) or percent change (Figure 3-11), depending on how
you chose to show the rates; then press Enter.
Up/Down Controller
Understanding and Adjusting Basal Rates
User Info/Support
During initial setup, you and your healthcare provider
either turned Off temporary basal rates or chose to
show the rates as % or U/hr. To change this setting, see
Chapter 6, Using the Personal Diabetes Manager.
Figure 3-10
4. Enter the duration for the temporary rate, then press Enter.
5. Press Confirm to start the temporary basal rate shown on the
screen. The pod beeps to indicate that the temporary basal
rate is running.
Blood Glucose Records
3
Figure 3-11
23