If The Pdm Becomes Submerged In Water - Insulet OmniPod User Manual

Insulin management system
Hide thumbs Also See for OmniPod:
Table of Contents

Advertisement

Appendices
2. Insert 2 new AAA alkaline batteries into the compartment.
The diagram inside the compartment shows you
which direction to insert the batteries (Figure A-3).
Figure A-3
3. Replace the battery door.
4. Turn the PDM face up.
5. Press the Power/Backlight
The PDM is now ready to use.
130
button.
Home

If the PDM becomes submerged in water

The PDM is not waterproof. Do not place it in water or leave it
near water where it can accidentally fall in. If it becomes sub-
merged in water:
1. Dry the outside of the PDM with a clean, lint-free cloth.
2. Open the battery compartment; remove the batteries and
discard them.
3. Use a clean, lint-free cloth to gently absorb any water in the
battery compartment.
4. Leave the battery compartment door open until the PDM is
thoroughly dry.
Never use a blow dryer or hot air to dry the pod or
PDM. Extreme heat can damage the electronics.
5. Do not put in fresh batteries or attempt to use the PDM until
it has thoroughly air-dried.
For 24-hour Customer Support, call 800-591-3455
(from outside the United States, 781-457-5098).