Integration Table - Simrad EY500 Instruction Manual

Portable scientific echo sounder
Table of Contents

Advertisement

Operation
The next column shows the total number of accepted single fish echoes found
within each layer. These numbers will give a good indication on the database for
the distribution that is given in the next 24 columns.
The numbers in the distribution field is the percentage occurrence in each TS
group. The total sum of all groups within each layer should add up to
approximately 100 %.
(7)

Integration table.

The layout for the integration results is shown below:
           
10.0 70
10
93/01/23 kHz
08.02.23
           
If the INTEGRATION TABLES parameter is enabled from the Printer Menu
and a log interval limit is reached, the integration results will be printed across
the paper. The format consists of 11 columns where the 10 rightwards columns
have identical layout. The left column carries the log distance, date, time and
transceiver frequency information. The other columns show the following
settings and results:
- Layer numbers (1 to 10)
- Layer type ( Surface, Bottom, Pelagic)
- Sv threshold
- Lower and upper layer limits referred to surface or bottom
- S
integration numbers (red colour). Note that the S
A
fish density calculations is S
- Average layer thickness (blue colour)
Note that the average layer thickness will become smaller when a layer is
intersected by bottom or transmitter pulse (The margin must also be included).
(8)
Event marker.
As for display echogram, except that Event Marker is controlled from the
Printer Menu. In addition the current event number and navigation data will
be written across the echogram paper.
14
9
8
7
6
5 Bot. -80
1.0 11.0
92 10.0
(mean) = S
/4B
a
A
4 Sur. -80
3 Sur. -80
2 Sur. -80
45.0 65.0
30.0 45.0
99 12.3
63 15.0
(mean) to be used for
a
P3402/C
1
Sur
-80
15.0 30.0
3.0 15.0
219 15.0
447 12.0

Advertisement