Simrad EY500 Instruction Manual page 221

Portable scientific echo sounder
Table of Contents

Advertisement

Simrad EK 500 / EY 500
Figure 6
Type F3. The display will show previously saved calibration files. The cursor
will appear at the end of the comment file on top of the display.
Erase the comment string and enter your comments concerning the present
calibration. In the Record Menu you have to fill in a new File Name and
Transceiver Number according to the setting in the sounder (if EK 500). The
transceiver number is selected by means of the horizontal arrow keys. The
correct TS for the calibration sphere according to actual sound velocity has to
be entered in addition to the Depth reading from the TS Menu provided that the
Transducer Depth has been set to 0.0 m. Refer to figure 7.
Figure 7
19
P2260/E