STEP1
STEP1: Dispose of used test strip.
STEP2: Remove wipe from the container and follow the instructions on the
package. If needed wring wipe slightly to remove excess liquid.
STEP3: Completely wipe the Glucose Meter including the front, back and sides,
and take care not to get any liquid in the test strip, test strip port and
serial port. Do not wrap the meter in a wipe. Take extreme care not to
get liquid in the test strip, test strip port and serial port of the meter. Allow
to remain wet for two (2) minutes at 68 .
47
STEP2
STEP3
℉
STEP4: Let the meter dry per the towelette manufacturer's instructions. Dispose
of wipe when finished.
User should wash hands thoroughly with soap and water after handling the
l
meter, or test strips.
The life of the device is three (3) years. The Element Compact meter was
l
demonstrated that there was no change in performance or in the external
materials of the meter after 11,000 cleanings with disinfection cycles
designed to simulate three (3) years for ten (10) times per day.
STEP4
48