Retesting Electrode Placement: Patient Sitting and Standing

1. Press the Stim tab and the L300 tab.
2. Press Test to turn on stimulation in the L300 only.
3. Press Stop to turn off stimulation.
4. If patient response is not accurate or is inconsistent with the original response, reposition
the L300 FS Cuff and assess the response to stimulation.

Attaching the L300 Cloth Electrodes and Bases

The L300 cloth electrodes are designed to provide an alternative for individuals who prefer to
not use the standard L300 hydrogel electrodes or have skin sensitivities (for example, allergy
or skin sensitivity to tape/adhesive). Fit standard L300 hydrogel electrodes first before fitting
the L300 cloth electrodes.
The L300 FS Cuff uses two L300 cloth electrodes and two L300 cloth electrode bases. See
Figure 8-20.
Figure 8-20: L300 cloth electrode and cloth electrode base.
Note: Wet the cloth electrodes before use. After every three to four hours of use, remove
and rewet the cloth electrodes. If the L300 FS Cuff is removed for more than one hour, then
wet the cloth electrodes again before use. If the L300 cloth electrodes dry out, the response
to the stimulation may change. If stimulation intensity needs to be adjusted more often than
usual, try rewetting the cloth electrodes.
CAUTION: The electrodes are for single patient use.
Faces Skin
Snaps to Base
L300 Cloth
Electrode
Chapter 8: Fitting the NESS L300 Plus System
L300 Cloth
Electrode Base
103

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NESS L300 Plus and is the answer not in the manual?

Table of Contents