4.4.2

Calibration menu

The calibration menu is used to adjust the sensors. The following functions are available:
Menu point:
1 Cy permeate
2 temp.permeate
3 calibr. reset
Service Handbook 08.2004
Function:
Calibration of permeate sensor. If the menu point '1 Cy permeate' is
chosen, a submenu appears with two items: '1 min. Cy' and '2 max. Cy'.
'1 min Cy': please remove the Cy sensor and press the . key.
'2 max Cy'
Important!!! The temperature must be calibrated before calibration of the
conductivity.
Connect the Cy sensor. Now the current permeate conductivity has to be
determined by means of a standard meter, and this value has to be
entered by numerical editing. The entered value is equated with the
current conductivity after actuation of the
Calibration of permeate temperature sensor.
For that purpose the current permeate temperature has to be defined
with a standard meter, and this value has to be entered by means of
numerical editing. The entered value is equated with the current
temperature when the
This menu point allows to reset adjustment of the two measuring inputs
to the internal standard values.
ATTENTION !! Internal calibration is not a correct adjustment.
key.
key is then actuated.