Malfunctions of the RO Control and Metering Pump

Indication
Malfunction
Indication of RO Fuse defective
control dark
Short circuit,
general
Solenoid valve
defective
Motor winding
defective
Water level in
Pump does not
container is
draw in
constant for
days
Pump does not
meter
Table 30
58
Cause
Corrective Action
Have replaced by
electrician
Replace defective parts
Replace defective parts
Replace defective parts
Suction line untight Retighten screwed
fittings. Replace
suction line
Suction line
Flush or replace
clogged
suction line
Crystals, debris in Clean valves
the valves
(metering head)
Membrane torn or
Replace membrane
membrane push-
rod torn out
Air in the suction
Bleed pump
line and in the
metering head
Lift adjustment a
While the pump is
zero
running, turn in "+"
direction (adjustment
values, see 3.3.1)
Crystals, debris in Clean valves
the valves
(metering head)
Valves not
Check correct line-up of
mounted correctly valves and correct,
if required (direction of
arrow!)
Injection fitting
Check direction of flow
blocked
and correct, if required,
or eliminate plug
(direction of arrow!)
WTC 600 LP / MP
Chap./
Fig.
ETL,
diagram
in switch
cabinet
ETL
ETL
ETL
4.2
4.2
4.2
Service Handbook 08.2004