The following table summarizes the high and medium priority alerts.
a
a
lErt
udIblE
I
ndICator
System Error
• •
• •
SmartCard
Error
5-2
t
ablE
v
Isual
I
ndICator
Backlights:
◊◊ ◊◊ ◊◊
Screen displays the
following message:
Service Required
Backlights:
Screen displays the
following flashing
error symbol:
d
a
p
EvICE
CtIon
ossIblE
The device enters the
Device failure
"Safe state" in which
the device power
remains on, but the
airflow is disabled.
The device continues
A problem exists
to operate and
with the Smart-
provide therapy, but
Card inserted in
data logging is un-
the accessory
available and FOSQ
slot. The card may
is deactivated.
be improperly in-
serted or the data
may be corrupt.
rEm
C
p
a
ausE
atIEnt
Press any of the display
screen buttons to silence
the alert. Remove the
power supply cord from
the device to remove
power. Plug the cord back
into the device's power
inlet to restore power. If the
alert continues to occur,
contact your home care
provider.
Remove the SmartCard
to clear the alert. Confirm
that the card is properly
oriented, with the arrow
pointing towards the de-
vice, as shown below, and
reinsert the SmartCard.
If the alert continues to
occur, remove the Smart-
Card from the device and
contact your home care
provider. The card may be
corrupt and need to be
replaced.
a
m s
u
star
uto
ErIEs
sEr
CtIon
m
anual