Intended Use • Warnings, cautions, and contraindications • System overview • Glossary and symbol key • How to contact Respironics ystEm ontEnts Your REMstar Auto M Series system includes the following items: Carrying Case User Manual Flexible Tubing Power Supply Reusable Gray Foam Filter Disposable Ultra-fine Filter AC Power Cord Device 1–1 s...
When in CPAP therapy, the system will deliver a continuous, set pressure during the night. Several accessories are available to make your OSA treatment with the REMstar Auto M Series system as convenient and comfortable as possible. To ensure that you receive the safe, effective therapy prescribed for you, use only Respironics accessories.
Page 7
• Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame. • When using oxygen with this system, a Respironics Pressure Valve must be placed in-line with the patient circuit. Failure to use the pressure valve could result in a fire hazard. • Do not use the device in the presence of a flammable anaesthetic mixture in combination with oxygen or air, or in the presence of nitrous oxide.
1.3.2 C autIons A Caution indicates the possibility of damage to the device. • The device may only be operated at temperatures between 41° F (5° C) and 95° F (35° C). • If this device has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature before starting therapy. Condensation may damage the device. • Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter. • A properly installed, undamaged reusable foam inlet filter is required for proper operation. • Tobacco smoke may cause tar build-up within the device, which may result in the device malfunctioning. Note: Additional warnings, cautions, and notes are located throughout this manual. 1.3.3 C ontraIndICatIons When assessing the relative risks and benefits of using this equipment, the clinician should...
The REMstar Auto M Series device, shown in Figure 1–2, is a CPAP (Continuous Positive Air- way Pressure) device designed for the treatment of Obstructive Sleep Apnea (OSA). It can deliver CPAP therapy or Auto-CPAP therapy. When prescribed for you, the device provides several special features to help make your therapy more comfortable.
Page 10
EvICE EaturE EsCrIptIon Accessory Module If applicable, insert the optional accessory card here. Air Outlet Connect the flexible tubing here. C-Flex Button This button allows you to change the C-Flex comfort setting, if C-Flex is prescribed for you. See Chapter 4 for more informa- tion about C-Flex.
Note: The exhalation port may be part of the interface or may be part of a separate exhalation device, but is required to minimize the potential for CO rebreathing. lossary The following terms and acronyms appear throughout this manual: Cronym EfInItIon Active State The state of the device when power is applied, the airflow is on, and the...
No User Serviceable Parts oW to ontaCt EspIronICs To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the Respironics Customer Service department at 1-800-345-6443 or 1-724-387-4000. You can also use the following address: Respironics...
Isplays This chapter describes the device’s control buttons and displays, patient circuit connections, and rear panel connections. ontrols and Isplays Figure 2–1 shows the three primary control buttons on the REMstar Auto M Series device. C-Flex 2–1 p IgurE rImary...
The screen times out after one minute if there is no activity and returns to the Standby screen. anEl Figure 2–4 shows the REMstar Auto M Series device’s rear panel (without a humidifier). DC Power Inlet Air Outlet Port...
3: s haptEr Etup This chapter provides instructions on how to: • Install the air filters • Position the device • Connect the breathing circuit • Supply power to the device nstallIng thE IltErs Caution: A properly installed, undamaged foam filter is required for proper operation. The device uses a gray foam filter that is washable and reusable, and an optional white ultra-fine filter that is disposable.
IrCuIt To use the system, you will need the following accessories in order to assemble the recommended circuit: • Respironics interface (e.g, nasal mask) with integrated exhalation port (or Respironics inter- face with a separate exhalation device such as the Whisper Swivel® II) • Respironics 6 ft. (1.83 m) flexible tubing • Respironics headgear (for the patient interface) Warning: If the device is used by multiple persons (e.g., rental devices), a low-...
Page 17
Connect the tubing to the mask: If you are using a mask with a built-in exhalation port, connect the mask’s connector to the flexible tubing, as shown in Figure 3–3. Exhalation Port Mask's Connector Flexible Tubing Connector 3–3 C IgurE onnECtIng a asK WIth uIlt...
upplyIng oWEr to thE EvICE You can power the device using AC or DC power. Caution: If this device has been exposed to either very hot or very cold tempera- tures, allow it to adjust to room temperature before beginning the following setup procedures.
3.4.2 u dC p sIng oWEr The Respironics DC Power Cord (Reorder Number 1001956) can be used to operate this device in a stationary recreational vehicle, boat, or motor home. The Respironics DC Battery Adapter Cable when used with the DC Power Cord, enables the device to be operated from a 12 VDC free-standing battery. Caution: When DC power is obtained from a vehicle battery, the device should not be used while the vehicle’s engine is running.
Page 20
omplEtE ssEmbly xamplE Figure 3–7 shows an example of how a complete assembly will look, with breathing circuit con- nected and power applied to the device. 3–7 f IgurE Inal ssEmbly xamplE hoWn WIth ptIonal CCEssory odulE Figure 3–8 shows an example of how you should route your tubing and situate your device on your nightstand for the best setup possible.
Page 21
4: d haptEr EvICE pEratIon This chapter explains the therapies available with the REMstar Auto M Series device, how to start the device, and how to change the settings. vaIlablE hErapIEs The REMstar Auto M Series device delivers the following therapies: •...
Page 22
Note: The FOSQ option will only appear on the Standby screen if a SmartCard is inserted in the device or if a modem is connected. 3. Press the button to turn on the airflow. The Active Display screen appears, shown in Figure 4–3.
sIng thE amp and EaturEs You can press the (Ramp) button to activate the Ramp feature, or the C-Flex button to change the C-Flex setting, if your provider has enabled these features. 4.3.1 r EaturE The device is equipped with an optional ramp feature that your home care provider can enable or disable.
4.3.2.2 IsablEd If C-Flex has not been prescribed for you and your provider has disabled the C-Flex feature, the following screen appears when you press the C-Flex button, and you cannot enable or adjust the setting. C-Flex Setting FLEX comfort has been disabled on this unit.
Page 25
4.4.1.1 IEWIng ata on thE atIEnt CrEEns Figure 4–6 shows how to navigate the Patient Data screens. Press Press – FOSQ Setup Therapy Usage Hrs/Night 7 Day Avg: 5:00 30 Day Avg: 8:00 Sessions > 4 7 Days: 5 Sessions 30 Days: 8 Sessions Apn-Hyp Index...
Page 26
Sessions Greater than 4 Hours Screen Sessions > 4 This screen displays the number of device therapy sessions that 7 Days: 5 Sessions exceeded 4 hours in 7-day and 30-day periods. The maximum value 30 Days: 8 Sessions that can be displayed for the 7-day period is 35 sessions, and the maximum value for the 30-day period is 150 sessions.
Page 27
Compliance Check Value Screen This screen displays a number used by your home care provider to validate that the data provided by you is the data taken from the Summary Data Screen. This is a “read-only” screen and you cannot change the number on the screen.
Page 28
The Patient Setup screens are described below. Standby Screen The Standby screen appears when you first apply power to the device. You can access the Patient Setup menu from this screen by pressing the + button. Once inside the Setup menu, press the –...
Page 29
Auto Off Screen Auto Off You can enable this feature if you want the device to automati- cally turn the airflow off whenever you remove the interface (e.g., Feature: ON mask) from your airway. Press the + or – buttons to select OFF or ON. If this feature is en- abled, the words Auto Off appear on the Standby screen if the de- vice detects that you have removed your mask.
2. From the Standby screen, press the – button to first FOSQ screen, shown in Figure 4–8. FOSQ 1/30 0 1 2 3 4 4–8 fosQ s IgurE CrEEn Press the + or – button to select your answer to the first question, and an arrow appears over your answer: –...
5: d haptEr EvICE lErts and roublEshootIng This chapter describes the device alerts and also provides troubleshooting information for issues you may run into when using the device. EvICE lErts The device provides two alert levels, high and medium priority. •...
Page 32
lErt ummary ablE The following table summarizes the high and medium priority alerts. lErt udIblE Isual EvICE CtIon ossIblE ausE atIEnt CtIon ndICator ndICator System Error • • • • Backlights: The device enters the Device failure Press any of the display ◊◊...
Page 33
lErt udIblE Isual EvICE CtIon ossIblE ausE atIEnt CtIon ndICator ndICator Mask Leak • • The device continues The breathing Check your breathing Backlights: to operate. circuit is discon- circuit connections and re- ◊ ◊ ◊ nected or there is connect the tubing if it has a large air leak.
roublEshootIng The table below lists some of the problems you may experience with your device or mask and pos- sible solutions to those problems. roblEm appEnEd hat to Nothing happens when There’s no power If you are using AC power, check the outlet you apply power to at the outlet or power and verify that the device is properly...
Page 35
roblEm appEnEd hat to The device’s display is The device or Unplug the device and the power supply. erratic. power supply has Reapply power to the device. If the problem been dropped continues, relocate the device to an area with or mishandled, lower EMI emissions (e.g., away from electronic or the device or...
Page 36
roblEm appEnEd hat to The device has fallen off The device may Always make sure your device is placed on a your table or nightstand. not have been hard, flat surface so the feet on the bottom of properly seated the device can adhere to the surface (make sure on the nightstand, there is no fabric under the device).
Page 37
roblEm appEnEd hat to You are having problems You have lost the If you are not using a humidifier, you cannot connecting the tubing to air outlet port or connect your tubing to the device without the the device. are traveling and detachable air outlet port, shown below.
Page 38
roblEm appEnEd hat to The air out of the mask The air filters may Clean or replace the air filters as described in is much warmer than be dirty. Chapter 7. usual. The device may be The temperature of the air may vary somewhat operating in direct based on your room temperature.
Some REMstar Auto M Series systems come with a SmartCard accessory module installed in the accessory slot on the back of the device to record information for the home care provider. Your home care provider may ask you to periodically remove the SmartCard and send it to him/her for evaluation.
• When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen. • When using oxygen with this system, a Respironics Pressure Valve must be placed in-line with the patient circuit. Failure to use the pressure valve could result in a fire hazard. • Oxygen accelerates fires. Keep the device and the oxygen container away from heat, open flames, any oily substance, or other sources of ignition. Do not smoke in the area near the device or the oxygen.
Allow the filter to air dry IltErs completely before reinstalling it. If the foam filter is torn, replace it. (Only Respironics-supplied filters should be used as replacement filters.) Filter Area If the white ultra-fine filter is dirty or torn, replace it.
Clean the tubing daily. Disconnect the flexible tubing from the device. Gently wash the tubing in a solution of warm water and a mild detergent. Rinse thoroughly. Air dry. ErvICE The REMstar Auto M Series device does not require routine servicing. Warning: If you notice unexplained changes in the performance of this device, if it is...
8: s haptEr pECIfICatIons nvIronmEntal pEratIng toragE 41° F (5° C) to 95° F (35° C) -4° F (-20° C) to 140° F (60° C) EmpEraturE 15 to 95% (non-condensing) 15 to 95% (non-condensing) ElatIvE umIdIty 77 to 101kPa (0 - 7500 ft) tmosphErIC rEssurE hysICal...
Page 44
rEssurE 4.0 to 20.0 cm H O (in 0.5 cm H O increments) Pressure Increments: Pressure Stability: <10.0 cm H O (±0.5 cm H ≥10.0 to 20.0 cm H O (±1.0 cm H Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15, and 20 BPM @ 20° C (±5° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals. Maximum Flow: 35 LPM Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15, and 20 BPM @ 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
Page 45
a: EmC I ppEndIx nformatIon – ’ uIdanCE and anufaCturEr EClaratIon lECtromagnEtIC mIssIons This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. mIssIons omplIanCE lECtromagnEtIC...
Page 46
’ – This device is intended for use in the uIdanCE and anufaCturEr EClaratIon lECtromagnEtIC mmunIty electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. IEC 60601 t mmunIty EvEl omplIanCE EvEl lECtromagnEtIC...
Page 47
(2) years from the date of sale by Respironics, Inc. to the dealer. If the product fails to perform in accordance with the product specifications, Respironics, Inc. will repair or replace – at its option – the defective material or part.