Permobil M300 Service Manual page 19

Hide thumbs Also See for M300:
Table of Contents

Advertisement

Drive wheels
Taking the rim apart
The rim can be taken apart to make it possible to fit/remove solid or pneumatic tires.
1. Remove the wheel in question from the wheelchair. See the previous page.
2. If the tire is pneumatic, release the air.
m WARNING!
Ensure that pneumatic tires are not pressurized before the rim is taken apart, otherwise there is a risk of personal injury.
3. Remove the six screws holding the two halves of the rim together (see illustration).
4. Take the rim apart.
Assembly
Assemble in the reverse order.
1. Fit the two rim halves (1&4) together with tire (3) and if pneumatic tire is used, it's inner tube (2).
Tighten the six screws using a torque wrench.
Tightening torque: 16.2 ft-Ibs (22Nm)
2. On wheels with pneumatic tires, fill the tire to recommended tire pressure, 29 psi (200 kPa).
3. Fit the wheel on to the wheelchair. See previous page.
m WARNING!
The recommended tire pressure for pneumatic tires is 29 psi (200 kPa). Overfilling entails a risk of explosion.
Incorrect tire pressure may result in lower stability and maneuverability. Check regularly that the tires have the correct pressure.
1
3
2
Fitting a tire to a split rim.
19
For this task the following tools are necessary:
1
Allen key 6 mm.
4
Pos.
1
Rim, outer section
2
Inner tube
3
Tire
4
Rim, outer section
5
Screw, ISO 4762 M8x30 8.8 Fe/Zn
Drive wheels
5
Description

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M400

Table of Contents