Replacing The Safety Belt; Testing An Actuator Motor - Invacare G50 Service Instructions Manual

Wheelchair
Hide thumbs Also See for G50:
Table of Contents

Advertisement

®
Invacare
- G50
7.20

Replacing the safety belt

Requirements:
• 10 mm socket spanner
• 4 mm Allen key
Dismantling the safety belt:
• Remove the plastic cap (5).
• Loosen the bolt (3) and the associated nut (in the figure this is covered) with a 4 mm Allen key
and a 10 mm socket spanner.
• Remove the nut incl. the washer.
• Remove the screw incl. the safety belt, the washer (2) and the washer arranged behind.
Note
Another nut is fixed between the two washers (2) and (4) as a spacer so that the belt mounting
can rotate freely.
• Replace the safety belt (1).
Refitting the safety belt:
• Refit the parts in reverse order.
7.21

Testing an actuator motor

Required parts/tools:
• Multimeter
• The electric resistance of the adjustment motor is to be checked at the connection assembly.
In case this is close to infinite, the motor is probably burnt out. If it is less than 1Ω, the motor
has short-circuited. In either case the motor must be replaced.
SERVICE INSTRUCTIONS
101

Advertisement

Table of Contents
loading

Table of Contents