Remote status bar
indicator LED blink
code
1
2
3
4
5
Servo light
module LED
blink code
ACS module defective
-
Plug on servo light module
1
(SLM) defective, or SLM
defective
Error at servo motor
2
Faulty accessory (e.g. short
-
circuit on adjustment motor)
Potentiometer on servo motor
3
(steering) defective
Lighting system defective
4
(error in SLM or in cabling)
Steering disengaged
5
Motor connection
-
loose/defective or motor
defective (M1)
Motor connection
-
loose/defective or motor
defective (M2)
Magnetic brake defective or
-
drive disengaged
SERVICE INSTRUCTIONS
Cause
Fault remedy
Replace faulty module
See "Replacing
electronics
components" on page 77
Replace faulty module
See "Replacing
electronics
components" on page 77
Replace servo motor.
Check accessory
connections, check
accessories See "Testing
an actuator motor" on
page 101
Replace potentiometer.
Check cable, replace
SLM. See "Replacing
electronics
components" on page 77
Re-engage the steering:
Switch the electronics
system off and on again.
Check the connecting
plug See "Checking the
cables" on page 87
Replace motor See
"Replacing a motor" on
page 41
Check the connecting
plug See "Checking the
cables" on page 87
Replace motor See
"Replacing a motor" on
page 41
Engage motor. Switch the
electronics system off and
on again.
Check the connecting
plug
Replace motor See
"Replacing a motor" on
page 41
23