Error Codes And Diagnostic Codes - Invacare G50 Service Instructions Manual

Wheelchair
Hide thumbs Also See for G50:
Table of Contents

Advertisement

SERVICE INSTRUCTIONS
6.1.3

Error Codes and Diagnostic Codes

6.1.3.1
CLAM
LED of the CLAM lights up constantly:
CLAM is functioning correctly.
LED of the CLAM is off:
CLAM may not be connected. Check plug-in
connections.
The joystick box might be programmed with an
incorrect driving programme (for example after
having replaced the joystick box). Check driving
programme.
LED of the CLAM is flashing:
CLAM is defective. Replace CLAM.
6.1.3.2
Joystick box / Servo-Light-Module
If the electronic system detects a fault this is indicated by a combined flashing code on the remote
and on the servo light module. Some faults can be remedied independently by the electronic
system. For this purpose, please switch the remote off and then on again. Please wait
approximately 5 seconds in each case before you switch the remote on again. If the fault can be
remedied, the flashing will stop on the status display. If the fault cannot be remedied by this,
please localise the fault by means of the following flashing codes.
Status display of
the remote (1)
22
Status display of the Servo-Light-Module (rear
cowling removed)
Invacare
Status display of the CLAM,
(visible through hole (1))
®
- G50

Advertisement

Table of Contents
loading