• Counter the dowel screw (the pivotal point of
the rocker) (1, not visible in the picture), using
the Allen key 6 mm to prevent it from rotating,
while using the open-end spanner 17 mm to
loosen and remove the nut (2).
• Pull the bolt out.
• Remove the trailing arm.
7.7.2

Assembly

• Replace the trailing arm with a new one
(assembly is done in reverse order to
disassembly).
• When inserting the lower fixation bolt of the
shock absorber, it can be helpful use a pair of
needle nose pliers to position the spacer
sleeves (1).
• Apply anaerobic metal adhesive to the
bushing and insert it into the trailing arm.
• Place the spherical disk on the axle (dowel
screw) with the flat side facing towards the
wheel.
• Assemble the wheel. Firmly tighten the
fixation nut.
SERVICE INSTRUCTIONS