Refrigeration Service Ports - Teledyne 4700 Serivce Manual

Refrigerated sampler
Hide thumbs Also See for 4700:
Table of Contents

Advertisement

Tools and supplies
Inspection
Service Port
(low pressure)
Figure 3-11 Refrigeration service ports
Refrigeration repairs are not performed by the factory, and
on-site repairs are not recommended. Teledyne Isco recommends
replacement of the refrigeration module in the event of com-
ponent failure.
If it is still deemed more practical to repair the system, on-site
repairs must be performed by licensed service personnel only.
Consult your local laws regarding refrigeration repair.
Copper Tubing
Tubing must be cleaned and capped copper,
refrigeration-grade. DO NOT use general pur-
pose tubing.
Tubing sizes:.250 O.D. x .190 I.D.
.312 O.D. x .280 I.D.
Saddle Tap Tool
For .312 O.D. service ports.
Silver Solder
Solder must be refrigeration-grade silver. DO
NOT use general purpose solder.
Lacquer Paint
Brazing and all exposed metal must be coated
Heat-Shrink
with lacquer paint to prevent corrosion.
Tubing
1. Using saddle tap tools (Figure 3-12), place pressure gauges
on both service ports (Figure 3-11).
2. Reconnect line power and allow the refrigerator to run,
observing the pressure gauges. Pressure on the high port
should be around 150 psi; pressure on the low port should
be around 10 psi.
4700 Refrigerated Sampler Service Guide
Section 3 Refrigeration System
Service Port
(high pressure)
3-9