roubleshooTIng and
When you have a problem or a question concerning what
you see on your screen, the following may be helpful.
What You See
42
Bayer Customer Service: 800-348-8100
What It Means
Test result is below
10 mg/dL.
Test result is above
600 mg/dL.
You have not pushed
the meter handle all
the way in.
• The 10-test disc is
out of test strips.
• There is no disc in
the meter.
• The disc has been
inserted upside down.
• The disc is defective.
The meter will mark
a test result with a
thermometer and an
if the meter is too cold
or too hot when the test
is performed. The result
will not be included in
any averages calculated.
e
C
rror
odes
What You Should Do
Repeat the test and make sure
the test strip is completely filled.
If LO appears again, contact
your physician or healthcare
professional.
Wash your hands and the
test site. Repeat the test. If
HI appears again, contact
your physician or healthcare
professional.
Push the meter handle in until it
clicks. You must push the handle
all the way in and see the "apply
blood" prompts before you touch
the strip to the blood drop.
Insert a new 10-test disc. If a full
disc is already inserted, remove and
inspect for foreign material. Rotate
the disc and reinstall it, bumpy side
up. If error repeats, try inserting
a new disc. If the error continues,
contact Bayer Customer Service.
Test only when the meter
temperature is between 50°F
and 113°F (10°C and 45°C).
: