Bayer HealthCare Breeze 2 User Manual page 50

Blood glucose monitoring system
Hide thumbs Also See for Breeze 2:
Table of Contents

Advertisement

What You See
Control Test
result is out of
range (too high
or too low).
Meter begins
countdown before
blood is applied.
Some display
segments do not
appear to be
showing when
is held in or
the meter handle
is all the way out.
46
Bayer Customer Service: 800-348-8100
What It Means
• Control solution is
past its expiration
date or is past the
6-month open
use-life date.
• 10-test disc is past its
expiration date.
• Deteriorated test
strip due to heat or
exposure to moisture.
• Control solution
is not at room
temperature.
• Not enough control
solution drawn into
test strip.
The inside of your
meter may be wet.
This may affect the way
you see your results.
What You Should Do
Run another control test. If the
result is still out of range, retest
with a new 10-test disc and
control solution with an acceptable
expiration date and open use-life
date. If still out of range, contact
Bayer Customer Service.
Wick the moisture away (see
page 36). If you continue to
experience difficulty, contact
Bayer Customer Service.
If there is a difference in your
meter's display and the one
shown on page 5, contact Bayer
Customer Service.