Error Codes and Symbols:
WHAT YOU SEE
24
Need Help? Call your local Bayer Diabetes Care Customer Service
WHAT IT MEANS
Low Battery.
Temperature
out-of-range.
The test strip has not
fi lled enough to give
an accurate result.
•
You may have
applied blood more
than once on the
same strip.
•
Blocked sample tip.
•
Blood drop too
small.
The meter is sensing
a used test strip.
WHAT YOU SHOULD DO
Replace the battery as soon as
possible (see page 28).
Move to an area that is within
the operating range for the
meter: 5°C–45°C. Let the meter
and test strips adjust to this
temperature for twenty minutes
before testing.
Remove the test strip and repeat
the test using a new test strip.
•
Do not attempt to apply blood
more than once per strip.
•
Do not push strip into skin
while fi lling strip.
•
See suggested drop size on
page 12.
Remove the test strip and retest
using a new test strip. Wait until
you see the fl ashing blood drop
in the display before adding the
blood sample.