Tilt-Foot Adjustment

The 6000 has front legs that extend to make it easier to view the LCD display. To
use them, place the 6000 on a stable base and rotate the legs downward until they
lock into position.
2.2.3

Rack Mounting

The 6000 may be rack mounted by using a 6000 rack mount kit. All rack mount
accessory kits contain detailed mounting instructions.
2.2.4

Ventilation Requirements

Rear panel area needs 2 to 4 inches of clearance for air circulation.
2.2.5

Power-Up Sequence

With the 6000 connected to an AC power source, set the power switch to "
supply power to the instrument and start the power-up sequence.
During the power-up sequence, the following takes place. For about 5 seconds an
initialization screen is displayed. The software version is displayed in the lower left
corner of the screen. During this time a self-test is performed to ensure that the 6000
hardware and software are communicating. If the 6000 cannot successfully complete
this test, an error message will be displayed.
After this test, the 6000 is configured to the state it was in when the power was last
shut off and displays the master display.
System Operation
WARNING
To avoid electrical shock hazard, connect the
instrument to properly earth-grounded, 3-prong
receptacles only. Failure to observe this precaution
can result in severe injury or death.