Chapter 4
4.2.3
Turn On Delay
The 6000 is CDRH Compliant with a user programmable turn on delay. The default
turn on delay is three seconds, but is user programmable from 0 to 30 seconds. The
delay setting is in the system configure screen, which can be reached by pressing the
MENU button, followed by the Config soft key, then the System soft key. The field
is labeled On Delay.
4.3
The Laser Connectors
On all laser modules, except the MOPA, a 9-pin female D-connector is used for input
and output connections, as shown by the pin-out diagram below. See Figure 12 for an
illustration of the back panel.
For MOPA modules, a 15-pin male D-connector is used for the MOPA cable and a 9-
pin male connector is used for the TEC jumper cable, as shown below.
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Laser Diode Driver Module Operation
Pin
Description
1,2
Interlock
3
Chassis Ground
4,5
Laser Cathode
6
Photodiode Cathode (+)
7
Photodiode Anode (-)
8,9
Laser Anode
Table 1 - Laser Connector Pinouts
MOPA Connector
Description
TE-
Analog Ground
OSC-/LAS-
OSC+
+12V Fan
LAS+
N/C
TE+
Interlock+
Therm-
Therm+
MPD+
MDP-
LED+
LED-
Table 2 - MOPA Laser Connector Pinouts
TEC Strap Connector
Pin
Description
1,2
TE+
5
Ground
6,7
TE-
8
Therm+
9
Therm-
41