Non-Verifiable Models; Verifiable Models - KERN GAB-P Series Operating Manual

Table scales
Table of Contents

Advertisement

6.8 Non-verifiable models:

 Switch on the balance and during the selftest press the
and the
„UnLoad" is displayed shortly.
 Ensure that there are no objects on the weighing pan.
 „LoAd" is displayed
 Place adjustment weight, the stability display
 After the adjustment the balance will carry out a self-test.
„PASS" will be displayed shortly. Remove adjusting weight
during selftest, the appliance will return into weighing mode
automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.

6.9 Verifiable models:

 Switch on balance and press the adjustment switch at the
same time.
 After that, during the balance's selftest, press
and
at the same time.
„UnLoad" is displayed.
 Acknowledge with
the weighing pan.
 The currently set adjustment weight will be displayed.
 To change by using the navigation buttons (see chap. 8)
select the desired setting, the active digit is flashing.
 Acknowledge with
18
button at the same time.
. Ensure that there are no objects on
. „LoAd" is displayed.
appears
GAB-P-BA-e-1613

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gab 6k-5pGab 10k-4pGab 30k-4p

Table of Contents