Permobil M300 Corpus HD Service Manual page 98

Hide thumbs Also See for M300 Corpus HD:
Table of Contents

Advertisement

Troubleshooting
Troubleshooting R-net
4.
Basic test
After a repair has been completed, the following test should be performed. These are minimum
recommendations. Depending on what the original error source was, further tests may be necessary.
m WARNING!
The tests described are minimum recommendations. It is the responsibility of the repair engineer(s) to perform other tests on the
basis of the original error source and the wheelchair model. The necessary information on other tests is available in the wheelchair
service manual. Permobil cannot be held responsible for losses of any kind that may arise when these tests are conducted, or that
arise as a consequence of further relevant tests not being conducted.
m WARNING!
These tests should be conducted in an open space, and some kind of clamping device, such as a safety belt, should always be
used. Permobil cannot be held responsible for losses for any kind arising due to these recommendations not being observed.
4.1
Basic inspection
Check that all contacts are properly connected.
Check all cables and contacts to ensure there is no visible damage.
Check that the rubber gaiter around the base of the joystick is not damaged. Inspect the gaiter
visually. It should not be subjected to manual handling.
Ensure that all components of the control system are securely installed.
Do not over-tighten the mounting screws.
98

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents