Removing/Installing The Pivot Tube; Removing - Invacare Pronto M50 Service Manual

Invacare pronto series power wheelchair service manual
Hide thumbs Also See for Pronto M50:
Table of Contents

Advertisement

Small Screw
Large Screw
FIGURE 10.4 Removing/Installing the Inner Shrouds

Removing/Installing the Pivot Tube

NOTE: For this procedure, refer to FIGURE 10.5 on page 92.

Removing

1. Remove the front headtube assembly. Refer to Removing/Installing the Front
Headtube Assembly on page 83.
2. Remove the side shroud. Refer to Removing/Installing the Side Shrouds on page 88.
3. remove the locknut, two small washers and large washer securing the rear of the pivot
tube to the threaded post.
4. Remove the pivot tube from the frame.
NOTE: The front and rear bushings may fall out of the pivot tube when it is removed. Retain these
for installation.
Part No. 1125075
SECTION 10—SHROUDS AND FRAME
Inner
Shroud
Large
Mounting
Hole
Small
Mounting
Hole
91
Side
Shroud
Walking
Beam
Pronto® Series