Motor/Gearbox/Brake - Invacare Pronto M50 Service Manual

Invacare pronto series power wheelchair service manual
Hide thumbs Also See for Pronto M50:
Table of Contents

Advertisement

8
E08
Controller Fault
9
E09
Communications
Fault
10
E10
General Fault
11
E11
Incompati-
ble/incorrect
Remote

Motor/Gearbox/Brake

SYMPTOM
Bad bearings.
Motor makes a
Raised commutator plate inside of
clicking noise.
motor.
Grinding noise or
Bad bearings. Bad Gears.
motor is locking
up.
Motors stall and
Current Rollback.
starts up again.
Part No. 1125075
SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
00
Controller fault
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing fault
09
Relay fault or precharge fault
10
Bridge fault or disable all fault
11
Electronics fault: Thermistor
12
Calibration setting fault
00
Remote connection lost
01
Low communication mode
00
General fault
00
Incompatible/incorrect
Remote
PROBABLE CAUSE
25
Check connections and wiring. Replace
power module.
Check connections and wiring. Replace
Bus cable.
Check all connections and wiring. Contact
Invacare Technical Service.
Wrong type of remote connected. Ensure
the branding of the joystick matches that
of controller unit.
SOLUTIONS
If bearings are bad, replace motor. Refer to
Removing/Installing the Motor on page 112.
Ohm out motor and replace motor if high reading
is present. Normal reading is .5-5 Ohms. Refer to
Motor Testing on page 31 or Removing/Installing
the Motor on page 112.
Replace motor/gearbox. Refer to
Removing/Installing the Motor on page 112.
Stop driving and let electronics cool.
Pronto® Series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pronto m51Pronto m61