2 of 8
The additions and changes to User Manual 34-SM-25-02 that relate to the newly
designed meter body and process heads are given in Table 1 of this addendum.
Use the information in Table 1 to reference and annotate your User Manual.
Table 1 Additions/Changes to the User Manual
Sub-Section
3.2 Considerations for SMV
3000 Transmitter
Table 3 Transmitter
Overpressure Ratings
4.5 Piping SMV 3000
Transmitter
Table 6 Installing ½ inch NPT
Flange Adapter
9.3 Inspecting and Cleaning
Barrier Diaphragms
Table 27 Inspecting and
Cleaning Barrier Diaphragms
9.5 Replacing Meter Body
Center Section
Table 29 Replacing Meter Body
Center Section
Replacement Parts
Figure 32 SMV 3000 Meter
Body
Table 38 Parts Identification for
Callouts in Figure 32
Wiring Diagrams and Installation
Drawings
34-SM-99-01 (Addendum to 33-SM-25-02)
Description of Change
The Maximum Working Pressure Rating and the
Overpressure Rating has been enhanced for all
models included in this addendum except for the
draft range transmitter.
For more information, refer to
Addendum
.
In Step 5 of Table 6, do not use the torque
specification of 47.5 to 54 N•m(35 to 40 lb-ft).
Instead, torque Flange Adapter bolts evenly to
47,5 N•m +/- 2,4 N•m (35 Lb-Ft +/- 1.8 Lb-Ft).
Do not use specifications for head bolt torque given
In Step 8 of Table 27.
Instead, torque head bolts/nuts to the specifications
given in Table 2 of this addendum.
Do not use specifications for head bolt torque given
In Step 9 of Table 29.
Instead, torque head bolts/nuts to the specifications
given in
of this addendum.
Table 2
Figure 32 illustrates and Table 38 lists the
replacement part available for the previous design
of the transmitter.
For the newer design, use Figure 1 of this addendum
to locate parts, and use Table 3 of this addendum for
part numbers and descriptions.
For applicability of parts, refer to
34-SM-03-01
SMV 3000 Smart Multivariable Flow Transmitter
Specification and Model Selection Guide
The numbers of installation drawings for transmitter
models of revision S and greater is given in Table 7
of this addendum.
in this
Table 1
03/04

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents