Honeywell SMV 3000 User Manual page 82

Smart multivariable transmitter
Hide thumbs Also See for SMV 3000:
Table of Contents

Advertisement

6.9
FlowConf Configuration - PV4,
PV4 (Flow) Upper
Range Limit (URL)
and Range Values
(LRV and URV)
ATTENTION
About URL and LRL
70
Set the URL, LRV, and URV for calculated flow rate PV4 output by
typing in the desired values on the FlowConf tab card.
URL = Type in the maximum range limit that is applicable for
your process conditions. (100,000 = default)
LRV = Type in the desired value (default = 0.0)
URV = Type in the desired value (default = URL)
Be sure that you set the PV4 Upper Range Limit (URL) to desired value
before you set PV4 range values. We suggest that you set the PV4 URL to
equal two times the maximum flow rate (2 x URV).
The Lower Range Limit (LRL) and Upper Range Limit (URL) identify the
minimum and maximum flow rates for the given PV4 calculation. The
LRL is fixed at zero to represent a no flow condition. The URL, like the
URV, depends on the calculated rate of flow that includes a scaling factor
as well as pressure and/or temperature compensation. It is expressed as the
maximum flow rate in the selected volumetric or mass flow engineering
units.
SMV 3000 Transmitter User's Manual
Continued
Continued on next page
1/99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents