8.2

Accessing Operation Data

Summary
Procedure
Table 25
92
You can access this data relevant to the operation of the transmitter using
the SCT.
• Current PV number selection
• Input
• Output
• Span
• Upper Range Limit
• Failsafe output direction
• Status
• Sensor (meter body) temperature
• Cold Junction Temperature
• High/low PV
• Lower Range Limit
• PROM serial number
• Scratch pad messages
Table 25 summarizes how to access the given operation data from the
transmitter using the SCT. The procedures assume that the SCT has been
connected and communications have been established with the transmitter
by selecting Tag ID menu item.
Accessing Transmitter Operation Data Using SCT
IF you want to view...
the present PV number
selected for display,
(transmitter in analog mode).
the status of transmitter
operation at the present time.
the PROM serial number.
the Firmware Version of the
transmitter.
the present message in the
scratch pad area of memory.
SMV 3000 Transmitter User's Manual
Select the SCT
Window or Tab Card
General Tab Card
Status Tab Card
Device Tab Card
And . . .
Read:
Analog Output
Selection
Read:
Gross Status
Detailed Status
Read: Serial Number
Firmware
Version
Scratch Pad
Continued on next page