  • Page 2: Product Description

    Product Description The Midas® S2 is a Fixed, Extractive Single-Point Gas Transmitter that draws a sample locally or from a remote point to a sensor cartridge. A wide range of toxic, flammable and oxygen gas sensor cartridges are available that enable the detection of gases used or generated in semiconductor manufacturing operations, light manufacturing environments, and national and university labs that specialize in semiconductor processing research.
  • Page 3 When Midas S2 reaches the end of its life, it should be disposed of by local regulations. Do not use cleaning solvents or abrasives to clean the gas detector.
  • Page 4: Hardware Overview

    Certifications Electrical: UL 61010-1, C22.2 No.61010-1, 61010-1, 61010-1 ROHS: 3(EU 2015/863) EMC: EN 50270:2015 What's in a full box System Midas® S2 transmitter Installation accessories (tubes, tube adaptor or LIT check valve if ordered) Quick reference guide ROHS sheet with drill template Hardware overview Display Pump door latch...
  • Page 5: User Interface Overview

    User Interface overview 1- Power 2- Alarm 1 3- Alarm 2 4 - Fault 5 - Tips for reset faults or alarms 8- Location label 6 - Buttons 7 - Device status icons 9 - Main status icons Cancel. Network Normal Reset connected...
  • Page 6: Specifications

    Specifications Transmitter Dimension Size (unit with Cartridge) 138.6 (H) x 55 (W) x 148 (D) mm (5.46 x 2.17 x 5.83 in) Weight - Transmitter 1.03kg (2.27lb) Weight - Sensor cartridge 34~73 g (0.075~0.161 lb) dependent on sensor type Power Requirements Operating Voltage 24 VDC Nominal, -15 to +10% (20.4 to 26.4 VDC)
  • Page 7: Cleaning The System

    Exhaust Length Up to 30 m (100 ft) Ambient Point In line air filter required Environmental Operating temperature 0°C to +40°C (32°F to 104°F) 0℃ to 30℃ (32℉to 86℉ ) with Pyrolyzer Condition Indoor only Operating atmospheric pressure 80~120Kpa Altitude <2000m Overvoltage Category 1...
  • Page 8: Mount The Device

    Mount the Device Mounting the device on a vertical flat surface. 1. Unscrew the two screws located on the main module. 2. Remove the main module by pulling it forwards off the base module. At this time, hold the cable gland in the base module with one hand, and with the other hand, carefully pull the main module.
  • Page 9 4. Drill two holes 2.3in (58.50mm) vertically apart for 2 x round head M4 screws, using the separately provided drill template on the ROHS sheet. 5. Partially screw the fixings into the mounting surface. 6. Place the mounting bracket assembly over the screws, so they pass through the mounting holes and then slide down to locate in the slots.
  • Page 10 8. Align the RJ45 socket of the power PCB module in the hole of the mounting bracket to refit the PCB. 9. Tighten the retaining screw to secure the power PCB module. 10. Wire the device as described in the "Wiring" section. 11.
  • Page 11 Pyrolyzer or NDIR connector Relay Contact terminals Midas S2 can be powered by either 24 VDC via traditional discrete wiring or by approximately 48 VDC delivered through the Ethernet cable from a PoE source. In either case, the 4-20 mA analog output can be used.
  • Page 12 NOTES: Midas S2 allows only one power source, 24 VDC or PoE. Failure to observe this requirement may cause damage to the gas detection system and will not be covered by the standard warranty Earthing Requirements: If the Midas S2 unit’s mounting bracket is not connected directly to a metal surface for earthing purposes, an additional earth wire will be required.
  • Page 13: Electrical Installation

    Electrical installation TB3-(1) TB3-(2) TB3-(3) TB3-(4) TB3-(5) TB3-(6) TB2-(2) TB2-(1) NDIR/ TB1-(2) PYRO TB1-(1) Reference Category Description TB1-(1) DC power 24Vdc input — VIN+ TB1-(2) DC power OVdc — VIN- TB2-(1) mA output 4-20mA analog output TB2-(2) mA output TB3-(1) Relay1 Alarm1 –...
  • Page 14 Note: Wiring should be executed within the red area.
  • Page 15 Attach the pyrolyzer or NDIR 1. Remove the main module as described in the Mount the Device process. 2. Plug the connector into the socket (as indicated in the image) at the bottom right of the terminal module through the rectangular access of the mounting bracket.
  • Page 16 Install the sensor cartridge 1. Turn the power off. 2. Verify that the part number and type of sensor cartridge are correct for your application, then remove the sensor cartridge from the packaging. Remove the BiAS battery module and plug caps from the sensor cartridge if needed.
  • Page 17 Turn the device on & off Line Integrity Test Honeywell offers the optional capability to check for leaks in Midas S2 sample lines. Midas S2 Transmitter for LIT (line Integrity Test) detects any leak in the tubing with a pneumatic signal change from the valves installed at the end of the sample line.
  • Page 18 Install and leak check 1. Mount the Midas S2 using the integral mounting bracket. 2. Connect the actual inlet & outlet tubes to Midas S2 and turn on the system. Verify the Check-Valve is attached to the Sampled Point. After the Power ON, wait approximately 10 minutes for pump stabilization and corrective measures.
  • Page 19 Tel: +41 (0)44 943 4380 (Alternative number) Middle East Tel: +971 4 450 5800 (Fixed Gas Detection) Middle East Tel: +971 4 450 5852 (Portable Gas Detection) Americas Honeywell Analytics Distribution Inc. 405 Barclay Blvd. Lincolnshire, IL 60069. USA Tel: +1 847 955 8200 Toll-free: +1 800 538 0363 detectgas@honeywell.com...
  • Page 20 RAE Systems by Honeywell Phone: 408.952.8200 Toll-Free: 1.888.723.4800 Asia Pacific Honeywell Co., Ltd. 5F SangAm IT Tower 434, World Cup buk-ro, Mapo-gu Seoul 03922 , Republic of Korea Tel : +82(0) 2 6909 0300 India Tel: +91 124 4752700 China Tel: +86 10 5885 8788 3000 analytics.ap@honeywell.com...
  • Page 21 Midas® S2 Manuals and other information about this product are available at : sps.honeywell.com Quick Reference Guide 3023M5000 Language: English Revision A @ 2023 Thursday, March 30, 2023 CoE CDMX...