Omega RD100B User Manual page 56

Programmable recorder communication interface
Hide thumbs Also See for RD100B:
Table of Contents

Advertisement

Example
Enable DST and set the DST start time to hour 0
on the 2nd Sunday of June and the DST end
time to hour 0 on the 2nd Sunday of December.
TD Use,Jun,2nd,Sun,0,Dec,2nd,Sun,0
Description • The DST start time and end time cannot be
set to the same time.
• The TD command can be used on the
RD100B and RD1800B with system version
1.11 or later. On the RD100B with system
version 1.02 or earlier, use the SS command.
SS
Sets the DST.
Syntax
SS p1,p2,p3<terminator>
p1 Standard time/DST selection
Winter
Standard time
Summer
DST
p2 Switch time setting ON/OFF (ON, OFF)
p3 Date and time (fixed to the YY/MM/DD HH
format)
YY
Year (00 to 99, the lower 2 digits)
MM
Month (01 to 12)
DD
Day (01 to 31)
HH
Hour (00 to 23)
Query
SS[ p1]?
Example
Switch to DST on 0 hour of June 1st, 2005.
SS Summer,ON,05/06/01 00
Description • Parameter p3 is not available when p2 is OFF.
• The SS command can be used on the
RD100B with system version 1.02 or earlier.
SK
Sets the computation constant
(/M1 option).
Syntax
SK p1,p2<terminator>
p1 Constant number (K01 to K30)
p2 Constant (up to 11 characters)
The range is –9.9999E+29 to –1.0000E–30,
0, 1.0000E–30 to 9.9999E+29.
(The + sign of "E+" can be omitted.)
SK[ p1]?
Query
Example
Set 1.0000E−10 to constant K01.
SK K01,1.0000E−10
Description This command cannot be specified while
computation is in progress.
SJ
Sets the timer used in TLOG
computation (/M1 option).
Syntax
SJ p1,p2,p3<terminator>
p1 Computation channel number
p2 Timer number (Periodic, 1, 2)
p3 Sum scale (for TLOG.SUM)
OFF
Integrates the data per scan interval.
/s
Integrates the physical amount that
are in unit of /s.
/min Integrates the physical amount that
are in unit of /min.
/h
Integrates the physical amount that
are in unit of /hour.
/day Integrates the physical amount that
are in unit of /day.
Query
SJ[ p1]?
Example
Enable timer 1 on computation channel 0B. No
sum scale designation.
SJ 0B,1,OFF
Description • This command cannot be specified while
computation is in progress.
• About p3
The data for sum computation are sampled
every scan interval. For data with units such
as /s, /min, /h, and /day as in a flow rate, the
data can be summed over the unit time as
shown below. Parameter p3 is valid only for
sum values.
OFF Σ (measured value)
/s
Σ (measured value) × scan interval
/min Σ (measured value) × scan interval/60
/h
Σ (measured value) × scan interval/
3600
/day Σ (measured value) × scan interval/
(3600×24)
The unit of the scan interval is seconds.
CM
Sets the communication input
data (/M1 option).
CM p1,p2<terminator>
Syntax
p1 Communication input data number (see
section 4.3)
p2 Communication input data
The selectable range is –9.9999E+29 to –
1.0000E–30, 0, 1.0000E–30 to 9.9999E+29.
(The + sign of "E+" can be omitted.)
Query
CM?
Example
Set 1.0000E–10 to communication input data
number C01.
CM C01,1.0000E−10
FR
Sets the interval for acquiring
data to the FIFO buffer
Syntax
FR p1<terminator>
p1 Pen model: 125ms, 250ms, 500ms, 1s, 2s,
2.5s, 5s, 10s
Dot model: 1s, 2s, 2.5s, 5s, 10s
FR?
Query
Example
Set the acquiring interval to the FIFO buffer to 1 s.
FR 1s
Description • Set the acquiring interval to an integer multiple
of the scan interval that is greater than the
scan interval.
• The scan interval on the pen model is 125 ms.
• The scan interval on the dot model varies
depends on the integration time of the A/D
converter as follows:
When the integration time is 16.7 ms or 20 ms;
6 channels: 1 s
12, 18, or 24 channels: 2.5 s