Performance Tests - Fluke 110 Calibration Information Manual

Table of Contents

Advertisement

Models 110, 111 & 112
Calibration Information
Function
(switch position)
B
B
+ Hz
D
s
e
R
S
J
(Models 111 & 112)
I
(Models 111 & 112)
1
Does not include test lead resistance.
12
Table 2. Performance Test Steps
Frequency
Input value
(sine wave)
300 mV
50 Hz
300 mV
500 Hz
5 V
50 Hz
5 V
500 Hz
50 V
500 Hz
600 V
500 Hz
5 V
50 kHz
0 V
5 V
0 Hz
-50 V
600 V
20
N/A
250
0
500
5 k
N/A
50 k
500 k
5 M
30 M
2 V
0 Hz
0.8 µF
N/A
9 A
500 Hz
5 A
0 Hz
-5 A
9 A
Display
Lower Limit
Upper Limit
0.294
0.306
0.294
0.306
4.947
5.053
4.947
5.053
49.47
50.53
593.7
606.3
49.93
50.07
-0.002
0.002
4.994
5.006
-49.94
-50.06
599.4
600.6
Beeper On
Beeper Off
0.2 1
-0.2
504.7 1
495.3
4.954
5.046
49.54
50.46
495.4
504.6
4.954
5.046
29.52
30.48
1.980
2.020
783 nF
817 nF
8.83
9.17
4.047
5.053
-4.047
-5.053
8.88
9.12

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

111112

Table of Contents