Required Equipment; Performance Tests; Test The Display - Fluke CNX i3000 Wireless iFlex Calibration Manual

Hide thumbs Also See for CNX i3000 Wireless iFlex:
Table of Contents

Advertisement

Required Equipment

The equipment in Table 2 is necessary for performance tests and calibration adjustment.
Equipment
Calibrator
Wired coil
Test Probe for iFlex
Test Lead

Performance Tests

To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the accuracy
of each function against Product specifications. If the Product fails a part of the test,
calibration adjustment and/or repair is necessary. See "Calibration Adjustment".

Test the Display

To verify that all segments of the display function:
1. With the Product off, push and hold .
2. Push  while you keep  pushed until all of the display segments are shown. See
Figure 1.
If segments of the display are missing, repair is necessary. See "Contact Fluke".