Fluke 113 Instruction Sheet

Fluke 113 Instruction Sheet

Electrical multimeter

Advertisement

Safety Information

A Warning statement identifies hazardous conditions and actions
that could cause bodily harm or death.
A Caution statement identifies conditions and actions that could
damage the Meter or the equipment under test.
To avoid possible electric shock or personal injury,
follow these guidelines:
Use the Meter only as specified in this instruction sheet
or the protection provided by the Meter might be
impaired.
Do not use the Meter or test leads if they appear
damaged, or if the Meter is not operating properly.
Always use proper terminals, switch position, and range
for measurements.
Verify the Meter's operation by measuring a known
voltage. If in doubt, have the Meter serviced.
Do not apply more than the rated voltage, as marked on
Meter, between terminals or between any terminal and
earth ground.
Use caution with voltages above 30 V ac rms, 42 V ac
peak, or 60 V dc. These voltages pose a shock hazard.
Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes,
or capacitance.
PN 3083192 December 2007, Rev. 1, 12/08
©2007, 2008 Fluke Corporation. All rights reserved.
Specifications subject to change without notice. Printed in China
Electrical Multimeter
Instruction Sheet
®
113

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 113

  • Page 1: Safety Information

    60 V dc. These voltages pose a shock hazard. • Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance. PN 3083192 December 2007, Rev. 1, 12/08 ©2007, 2008 Fluke Corporation. All rights reserved. Specifications subject to change without notice. Printed in China...
  • Page 2 • Do not use the Meter around explosive gas, vapor or in damp or wet environments. • When using test leads or probes, keep your fingers behind the finger guards. • Only use test leads that have the same voltage, category, and amperage ratings as the meter and that have been approved by a safety agency.
  • Page 3 Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycle information. Terminals LOW IMPEDANCE 600 V 300 V fda09.eps Description Common (return) terminal for all measurements.
  • Page 4: Automatic Range Selection

    Automatic Range Selection The meter defaults to autoranging when turned on. To return to autoranging while in the manual range mode, press qfor 1 second. Manual Range Selection ( k CHEK, Ω and E See each function in this instruction sheet for available ranges. RANGE fda01.eps...
  • Page 5: Battery Saver ™ (Sleep Mode)

    ™ Battery Saver (Sleep Mode) If the Meter is ON, but inactive and not connected to voltage for more than 20 minutes, the display goes blank to preserve battery life. To resume operation, press any button or turn the rotary switch.
  • Page 6: Power-Up Options

    Power-Up Options To select a Power-Up Option, hold down the button indicated in the following table while turning the Meter from off to the k CHEK function. Power-Up Options are canceled when you turn the Meter off and when sleep mode is activated. Button Power-Up Options Turns on all display segments until the button is released.
  • Page 7: Ac And Dc Voltage

    W k CHEK If a dc or ac voltage greater than about 3 V is present across the inputs when the meter is set to k CHEK, the meter switches automatically to dc or ac voltage mode and displays the voltage. When k CHEK is activated, the meter has low input impedance (LoZ) ≅3 kΩ.
  • Page 8: Diode Test G

    Diode Test G Turn off circuit power before testing. For best results, diodes should be measured out of circuit. Also refer to k CHEK Good Diode Good Diode Good Diode Single Beep Reverse Bias Forward Bias fda03.eps Bad Diode Bad Diode Bad Diode Open Shorted...
  • Page 9: Continuity Capture

    Continuity s Turn off circuit power before testing. Open Short fda05.eps Continuity Capture Turn off circuit power before testing. MIN MAX fda06.eps...
  • Page 10 Short Open Short to Open Open to Short fda10.eps Captures transitions longer than 500 μs (1/2000 of a second). Transitions after the first transition cause the Meter to beep, but the display does not change. To reset the display to the current condition, press p. To exit, press pfor 2 seconds, or turn the rotary switch.
  • Page 11 Resistance e Turn off circuit power before testing. RANGE fda07.eps Capacitance E Turn off circuit power, then disconnect and discharge the capacitor before measuring capacitance. RANGE fda08.eps Note correct probe polarity for polarized capacitors.
  • Page 12 MIN MAX (Records the lowest, highest, and calculated average measurements.) Autoranging, and Battery Saver™ are disabled. Put the meter in the proper range before entering MIN MAX. When a new minimum or maximum is recorded, the meter beeps. Enter MIN MAX Maximum Reading Minimum Reading 2 sec...
  • Page 13: Disabling The Beeper

    This meter should be serviced only by a qualified service technician. To locate an authorized service center, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www.fluke.com.
  • Page 14: Specifications

    Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 95 %. Accuracy specifications are given as follows: Note Specifications are subject to change without notice. ±([% of reading] + [number of least significant digits]) Accuracy Function...
  • Page 15 Maximum Voltage Between any Terminal 600 V and Earth Ground: Display: 3 3/4-digits, 6000 counts, updates 4/sec Operating Temperature: -10°C to 50°C (14°F to 122°F) Storage Temperature: -40°C to 60°C (-22°F to 140°F) 0.1 x (specified accuracy)/°C (<18 °C or Temperature >28 °C) Coefficient:...
  • Page 16 Function Input Impedance (Nominal) k Chek ~3 kΩ <300 pF Common Mode Rejection Ratio (1 kΩ Unbalanced) k Chek >60 dB at dc, 50 or 60 Hz Open Circuit Test Voltage Full Scale Voltage <2.7 V dc <0.7 V dc Ω...
  • Page 17 LIMITED WARRANTY & LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.