Performance Tests; Required Equipment - Fluke 110 Calibration Information Manual

True-rms multimeter
Table of Contents

Advertisement

110/113/114/115/116/117
Calibration Information

Performance Tests

To prevent electric shock, do not perform the performance test procedures unless the Meter is
fully assembled.
The following performance tests verify the complete operation of the Meter and check the accuracy of each
Meter function against its specifications. The recommended calibration interval is 12 months. If the Meter fails
any part of the test, calibration adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the device under test (DUT).
See Table 1 for a list of the features for each Meter.

Required Equipment

Table 2 is a list of the equipment required to conduct a performance test on the Meter.
Equipment
5522 Multi-product Calibrator (or equivalent)
Fluke 80 AK
K-type Thermocouple Adapter Accessory
K-type Thermocouple, mini-plug on both ends
Double Banana plug
[1] For accuracy specifications, see the Users Manual for the product.
6
XW Warning
Table 2. Equipment Requirements
Measurement Function
DC Volts
DC Current
AC Volts
AC Current
Resistance
Capacitance
Temperature
Frequency
Temperature
Temperature
VoltAlert
Range
[1]
10 mV to 600 V
600 μA to 10 A
6 mV to 600 V
600 μA to 10 A
Ω
0 to 30 M
9 to 900 μF
0 °C to 400 °C
2 V, 50 kHz
NA
NA
NA