13.5. Dust and Drip Proof
IEC IP x2
Water Prevention
Level
Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up
2
to 15° from its normal position.
13.6. Cable Strength
There was no sign of malfunction with the scanner's structure and its performance after
the following static load test: With the scanner fixed, pulled the cable with static load 24.5
N (or 2.5 kgf) for a second. Repeated this routine 20 times.
13.7. Cable Bending Test
There was no sign of malfunction with the scanner's structure and its performance after
the following flexing test: Applied load 4.9N (or 500gf) to the cable and flex it 60 degrees
to the left and right as shown below. Repeated this routine one million times.
Note: Cable bending resistance is not warrantable.
Figure 27: Cable bending test
Details
4 .9 N (500g f )
4 .9 N (500g f )
35
Opticon
OPL 7836
Specifications Manual