Item
100 µL piston kit
250 µL piston kit
500 µL piston kit
1 mL piston kit
5 mL piston kit
10 mL piston kit
25 mL piston kit
6-6
Changing the Syringe Piston
Part Number
Assembly
F4015072
F4015073
For all syringe sizes, it is necessary to replace the
F4015074
piston assembly after approximately 30 000 cycles
F4015075
(using water).
F4015076
The correct reference piston kit must be ordered:
F4015077
F4015078

Replacing the Piston

1) Dismount the Syringe assembly from the instru-
ment, as described page 3-9.
2) Pull the piston out of the syringe body.
3) Reassemble the syringe and piston as described
page 6-7.
4) Remount the syringe assembly on the valve or
Tee junction.