Dilutor dispenser

Advertisement

Installation
3-6
The transfer tubing connects the outlet port (right-
hand) of the Tee-junction to the probe. A 2 mL
tapered transfer-tubing is supplied as standard ;
1 mL, 5 mL, and 10 mL are available as additional
accessories.
Self-adhesive clips are provided for holding the
transfer tubing securely against the side of the 402
(see previous page).
The 1 mL transfer tubing is recommended for
transferring volumes smaller than 1 mL. To avoid
movements of the tubing that could affect the
precision, the 5 mL and 10 mL transfer tubings are
rolled and held by magnets on the left-side of the
402.
After assembling the probe, the other end of the
tubing is installed by screwing the connector into
the right-hand port of the Tee-junction. When you
use Diamond tips, the special non-tapered tubing,
included in the kit, must be fitted.