Entering The Configuration (Conf) - Gilson 402 User Manual

Dilutor dispenser

Advertisement

Operation - Advanced Use
Configuration Screen
Tubing Volume Screen
5-6
4 Press the Next soft-key to go to the following
screen.
5 Press the Chge soft-key to choose an option or to
select from a list of possible values.
6 Press the ESC key to return to the level of the
previous screen or menu.
7 Press CLEAR to remove the value (before
pressing ENTER) and return to the previous
value.
8 Press Deflt to select the default value or option.

Entering the Configuration (CONF)

You have to tell the controlling software how your
402 Dilutor-Dispenser is configured (syringe sizes,
flow rates, etc.). Press the CONF key; you see the
Configuration Screen.
Press Key To Change
Vol
Volumes
Syr
Syringe
Flow
Flow-rates
Count
Counters
Etc
Switch On/Off
Volume(s)
Press the Vol soft-key, you see the Tubing Volume
Screen.
The use of the soft-keys is as follows:
Next is used to go to the next screen (from Tubing
Volume Screen),
Chge is used to select the tubing size,
Prev is used to go to the previous screen (from Air
Gap Volume Screen),
Check that the configured volume for the probe
tubing is the same as the size fitted. For the
standard probe, you can select from: 1, 2, 5, 10 mL
(tapered tubing).
When using disposable tips, you must select the size
of tip (Micro or Large Volume), and application
type for Micro Tips only (Air or Liquid). When you
use disposable tips, the 'Extra Volume' parameter is
automatically changed to 'Air Push Volume'.
What Items
Transfer Tubing, Air Gap.
Sizes, Correction, and to change.
Air, Prime, Reservoir.
Number of cycles for syringe(s) and valve.
Beeper and Autorefill function.