Error Messages
Error Message
Parameters incompatible!
Ratio impossible
Ratio missing
Reservoir not allowed
Right syr life exhausted!
Sample too small
Sample volume missing
Syr capacity too small
Syringe Life Exhausted!
Total volume is missing!
Total volume missing
Total volume too small
Valve life exhausted!
xxx exists, overwrite?
Cause/Possible Action
Refer to page 5-16.
Dilution ratio calculated by software is not valid; check
parameters using formulae given in page 5-18 (DILUTE task)
and correct one or both of the other values.
Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Liquids must be aspirated using the probe.
For the Dual syringe model: piston or right-hand syringe may
need servicing (refer to page 6-6).
Enter a larger value for Sample Volume. Verify parameters in
the task.
Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Check volumes used in the task correspond to the configured
syringe size(s).
For the Single syringe model: piston syringe may need
servicing (refer to page 6-6).
Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Re-enter Method, supplying the missing parameter, when
prompted to do so (see DILUTE task).
Enter a larger value for the Total Volume, or check the volumes
used in the task.
Valve may need cleaning or servicing (refer to page 6-2).
Select another file number, unless you want to overwrite file xxx.