Dilutor dispenser

Advertisement

Operation - Advanced Use
File Number
Method 90
Method 91
5-28
Key in the file number of the second Method to
link(8), press ENTER, press Next.
Key in the file number of the third Method to
link(9), press ENTER, press Next.
Key in the file number of the fourth Method to
link(8), press ENTER, press Next.
After a file is linked in this manner it can not be
deleted.
Press ESC to access the Save Screen, press Yes.
Typical Task Entry
Key to Press:
File # (44) ENTER
Prefabricated Methods
Five prefabricated Methods, to run or to adapt, are
present in the 402's memory as file numbers 90 to 94.
Although the tasks are the same, the parameters
vary according to whether you have a Single-
syringe or a Dual-syringe 402.
Configuration
Standard equipment with a 2 mL transfer tubing.
Left-hand (or only) syringe: 5 mL.
Right-hand syringe (if fitted): 500 µL.
Default values are used for parameters not listed.
TASK/Parameter
Dilute Normal
Dilution Ratio
Sample Volume
Rinse
Dilute Serially
Total Volume
Sample Volume
Nb of Dilutions
Value for use with
Dual-syringe Model
10
200
300
3000
100
3
Edit
More
Link
File # (7) ENTER
Next
File # (8) ENTER
Etc until all file numbers
have been entered then:
ESC
Single-syringe Model
10
1000
1500
3000
1000
3
5