Download Print this page

Gilson 402 User Manual page 112

Dilutor dispenser

Advertisement

Technical Data
For 1 % syringe capacity, typical
values are given.
Appendix C-4
The same specifications are obtained in the
dispensing mode with liquid from the reservoir.
See OQ Document for procedure and calculations.
Syringe
Volume
capacity (µL) pipetted (µL) %
100
1
250
2.5
500
5
1000
10
5000
50
10000
100
25000
250
Typical values (all syringes)
Syringe capacity Max. accuracy Max.error imprecision
%
100
50
10
Junction Tubing
(between valve and Tee)
Material: FEP, with two PVDF union connectors.
Tubing dimensions: 2.0 mm i.d., 70 mm length.
Tubing dead volume: 220 µL.
Appendix
Accuracy
µL
5.0
0.05
3.0
0.075
3.0
0.15
3.0
0.3
3.0
1.5
3.0
3
3.0
7.5
+ A %
RSD %
0.50
0.10
0.80
0.20
1.00
0.50
C
RSD
%
µL
3.0
0.03
2.0
0.05
2.0
0.1
1.0
0.1
0.8
0.4
0.5
0.5
0.5
1.25

Hide quick links:

Advertisement

loading
Save PDF