Equipped X-ray Generator
06/2009
Parameters
IEC standard 60601-2-28 (1993)
Type of protection against electric shocks
Degree of protection against electric shocks
Rated value of inherent filtration
Rated value of additional filtration
Rated value of total filtration
Beam-limiting cone, focal spot/skin distance
Maximum accumulated heat.
Maximum continuous thermal dissipation
Amount of leaking radiation at maximum rate
during one hour of use
Maximum field of symmetrical radiation
Position and tolerance of the focal point on
the reference axis
a) Reference axis
b) Target angle
c) Focal point
SM731_K2200_01_en
Table 6: Equipped x-ray generator
Figure 14: X-ray generator
Features
Conform
Class 1
Type B
1.5 mm (0.059 in.) eq. Al
1.0 mm (0.039 in.) eq. Al
2.5 mm (0.098 in.) eq. Al
20 cm (7 7/8 in.)
32,500 J
7 W
< 0.25 mGy
6 cm (2 3/8 in.) diameter
0 mm +/- 0.5 mm (0.020 in.)
6-5