Adjusting T-Arm Sockets - Invacare ProSPIN X4 Owner's Manual

Table of Contents

Advertisement

SECTION 8—ARMS

Adjusting T-Arm Sockets

NOTE: For this procedure, refer to 
FIGURE 8.7.
NOTE: Perform this procedure if the T‐arm is 
too loose in the socket or does not easily slide up 
and down in the socket.
1. Remove the rear wheels from the 
wheelchair, if necessary. Refer to 
Removing/Installing Rear Wheels on 
page 67.
2. Loosen, but DO NOT remove the four 
hex screws and washers that secure T‐
arm socket to T‐arm clamp.
NOTE: The T‐arm socket will disassemble if 
the four hex screws and washers are removed.
3. Slide the T‐arm into the T‐arm socket until the locking lever is in the slot in the T‐arm 
socket and an audible ʺclickʺ is heard.
4. Squeeze the T‐arm socket and the T‐arm clamp together until the socket is flush with 
the T‐arm.
5. While holding the T‐arm socket and the T‐arm clamp together, tighten the four hex 
screws and washers. Tighten securely.
6. Press in on the locking lever and lift the T‐arm straight up and out of the T‐arm socket.
7. Repeat STEPS 3‐6, if necessary until the T‐arm slides in the T‐arm socket as desired.
8. If necessary, install rear wheels. Refer to Removing/Installing Rear Wheels on page 67.
ProSPIN™ X4™
T-Arm
Hex Screws
and Washers
Slot
Locking Lever
FIGURE 8.7 Adjusting T-Arm Sockets
52
T-Arm Clamp
Hex Screws
and Washers
T-Arm Socket
Part No 1125104