SECTION 7—FRONT RIGGINGS

Adjusting Adjustable Angle Flip-up Footplates

WARNING
When determining the angle of the footplates, make sure the rear of the footplates
DO NOT interfere with the movement of the front casters. Otherwise, injury or
damage may occur.
NOTE: For this procedure, refer to FIGURE 7.3.

Adjusting Depth

1. Remove the two flat screws and locknuts that secure footplate to the half clamp.
NOTE: Observe the angle of the footplate for reinstallation.
2. Move the footplate to one of four mounting positions.
NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180°.
3. Tighten the two flat screws and locknuts.
NOTE: The setting for positioning the footplate on the half clamp may vary for each footplate.
4. Repeat STEPS 1‐3 for opposite footplate, if desired.

Adjusting Angle

1. Loosen, but DO NOT remove, the adjustment screw in the half clamp.
2. Position the footplate to the necessary angle to accommodate the user (Detail ʺAʺ of 
FIGURE 7.3).
3. Retighten the adjustment screw. Torque to over 90 in‐lbs, but no more than 300 in‐lbs.
4. Repeat STEPS 1‐3 for opposite footplate, if desired.
90° Footrest Support
DETAIL "A" - SIDE VIEW OF
FOOTPLATE AND
FOOTREST SUPPORT
Flat Screws
Footrest Support
Footplate
Footplace
Footplace
Half
Hinge
Clamp
Adjustment Screw
Locknuts
FIGURE 7.3 Adjusting Adjustable Angle Flip-up Footplates
42
ProSPIN™ X4™
Part No 1125104