Cleaning The Tubing; Cleaning The Water Chamber; Service - Respironics SleepEasy User Manual

Sleepeasy with c-flex
Table of Contents

Advertisement

1. If the device is operating, stop the airflow by pressing the
2. Remove the filter(s) from the enclosure by gently squeezing the filter in the center and pulling it away from the
device.
3. Examine the filter(s) for cleanliness and integrity.
4. Wash the gray foam filter in warm water with a mild detergent. Rinse thoroughly to remove all detergent residue.
Allow the filter to air dry completely before reinstalling it. If the foam filter is torn, replace it. (Only Respironics-
supplied filters should be used as replacement filters.)
5. If the white ultra-fine filter is dirty or torn, replace it.
6. Reinstall the filters, inserting the white ultra-fine filter first if applicable, refer to "Installing the Air Filters".
CAUTION: Never install a wet filter into the device. You must ensure sufficient drying time for the cleaned filter.

Cleaning the Tubing

Clean the tubing daily. Disconnect the flexible tubing from the device. Gently wash the tubing in a solution of warm
water and a mild detergent. Rinse thoroughly. Air dry.

Cleaning the Water Chamber

Note: The Flapper (Soft) Valve has a one year lifespan. You must replace this valve after one year.
Note: Hand washing can be performed daily. Dishwashing (water chamber and base plate only) can be performed
once a week. The flapper valve and protective cover must be hand washed.
WARNING: Empty and clean the water chamber daily to prevent mold and bacteria growth.
WARNING: Allow the water in the chamber to cool to room temperature before removing the chamber from
the humidifier.
1. Turn the therapy device off and allow approximately 15 minutes for the heater plate and water to cool.
2. Disconnect the tubing from the water chamber.
3. Slide the water chamber out of the humidifier platform. Empty any remaining water.
4. Gently remove the base of the chamber with your hands, being careful not to damage the rubber seal.
5. Remove the flapper valve from the air inlet port; unclip the protective cover from the top of the water chamber.
Wash these two parts by hand only in a solution of warm water and mild dishwashing soap. Rinse with clean water
and allow them to air dry.
6. Wash the water chamber and base plate in the dishwasher (top shelf only) or by hand in a solution of warm water
and mild dishwashing soap.
7. Inspect all parts for damage prior to reassembly.
8. Reassemble the water chamber. Make sure the base plate is fully seated on the water chamber. Also, make sure
you install the flapper valve properly into the air inlet port of the water chamber.
Note: Be sure to reinstall the protective cover on top of the water chamber to prevent debris from entering the
water chamber.
9. Fill the water chamber to the fill line. Inspect the water chamber for any leaks or damage. Replace the water
chamber if any damage is present.

Service

The SleepEasy device does not require routine servicing.
WARNING: If you notice unexplained changes in the performance of this device, if it is making unusual or harsh
sounds, if the enclosure is broken, or if water has entered the device, discontinue use, and contact your home care
provider.