3. Before each use, rinse the chamber with water. Fill the chamber to the fill line with distilled water using the funnel.
CAUTION: Use only distilled water in the chamber. The use of additives or chemicals in the chamber is
prohibited.
CAUTION: Do not fill water chamber above the fill line indicated on the side of the water chamber. Damage to
the therapy device may occur.
CAUTION: Do not allow the water chamber to sit for any length of time after it has been filled with water.
Immediately install the chamber in the device. Allowing water to sit in the chamber (when the chamber is not
installed in the device) may cause the chamber to separate from the bottom plate and may cause water leakage.
4. Slide the chamber into place on the side of the unit.
WARNING: Never touch the heater plate unless the unit is unplugged and the plate has cooled down.
CAUTION: Avoid moving the device when the water chamber has water in it.
5. Connect the flexible tubing (included with your therapy device system) to the outlet port on the water chamber.
IMPORTANT: Before each use, examine the flexible tubing for any kinks, damage, or debris. If necessary, clean
the tubing to remove the debris. Replace any damaged tubing.
6. The ideal humidity setting depends on room temperature and humidity. Initially, a setting of 2 is recommended.
You can adjust this setting at any time. Press and hold the humidifier button on the therapy device. The humidifier
symbol and setting will appear. Press the + or - buttons to change the setting.
IMPORTANT: When the airflow is turned off, the humidifier will automatically shut off. If you restart the airflow,
the heated humidification will return to the previous setting. Loss of power to the system will require the user to
press the heated humidifier button to reactivate the humidifier.

Display

The display screen is shown here.
The information shown on the display screen is defined as follows:
i
con
Indicates that the Blower Hours Time Meter is
being displayed.
Indicates that the Therapy Hours Time Meter is
being displayed.
When displayed with the Therapy Hours icon,
it indicates that the Session Counter is being
displayed.
Indicates that the humidifier is providing heat.
Indicates that the unit is in the Provider Settings
Menu.
In Active State, indicates that the Ramp function is in
progress. In menus, it is used to indicate Ramp Time
Setting alone, or in combination with Ramp Start
Pressure.
U
M
ser
anUal
D
eScription
i
con
Indicates that the Ramp Start Pressure is being
displayed.
Indicates that a pressure value is being displayed.
Indicates that the user may erase the displayed
data.
Indicates that the unit requires user attention.
Indicates that the Compliance Check Value for the
unit is being displayed.
Indicates that the C-Flex setting is being displayed.
D
eScription
9